Calibration of RADMON radiation monitor onboard Aalto-1 CubeSat
Praks J; Gieseler J; Dudnik O; Hæggström E; Hedman H; Punkkinen R; Oleynik P; Säntti T; Niemelä P; Valtonen E; Punkkinen A; Peltonen J; Vainio R; Hietala H
Calibration of RADMON radiation monitor onboard Aalto-1 CubeSat
Praks J
Gieseler J
Dudnik O
Hæggström E
Hedman H
Punkkinen R
Oleynik P
Säntti T
Niemelä P
Valtonen E
Punkkinen A
Peltonen J
Vainio R
Hietala H
Elsevier Ltd
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2021042612709
https://urn.fi/URN:NBN:fi-fe2021042612709
Tiivistelmä
RADMON is a small radiation monitor designed and assembled by students of University of Turku and University of Helsinki. It is flown on-board Aalto-1, a 3-unit CubeSat in low Earth orbit at about 500 km altitude. The detector unit of the instrument consists of two detectors, a Si solid-state detector and a CsI(Tl) scintillator, and utilizes the ΔE-E technique to determine the total energy and species of each particle hitting the detector. We present the results of the on-ground and in-flight calibration campaigns of the instrument, as well as the characterization of its response through extensive simulations within the Geant4 framework. The overall energy calibration margin achieved is about 5%. The full instrument response to protons and electrons is presented and the issue of proton contamination of the electron channels is quantified and discussed.