Contemporary radiation doses in interventional cardiology: a nationwide study of patient doses in Finland

Markku Pirinen; Teemu Siiskonen; Hanna Matikka; Joanna Sierpowska; Jukka Järvinen; Teuvo Parviainen; Saija Hurme; Tuomas T Rissanen; Heli R S Larjava; Eini Niskanen; Timo J Mäkelä; Satu Strengell; Mika Teräs; Markku Eskola; Tuomas Kiviniemi; Hannu Järvinen; Antti Kivelä; Elina Hallinen

Contemporary radiation doses in interventional cardiology: a nationwide study of patient doses in Finland

Markku Pirinen
Teemu Siiskonen
Hanna Matikka
Joanna Sierpowska
Jukka Järvinen
Teuvo Parviainen
Saija Hurme
Tuomas T Rissanen
Heli R S Larjava
Eini Niskanen
Timo J Mäkelä
Satu Strengell
Mika Teräs
Markku Eskola
Tuomas Kiviniemi
Hannu Järvinen
Antti Kivelä
Elina Hallinen
Katso/Avaa
Final draft (post-print) (843.0Kb)
Lataukset: 

Oxford University Press
doi:10.1093/rpd/ncz041
URI
https://academic.oup.com/rpd/article/185/4/483/5466141
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2021042824431
Tiivistelmä

The amount of interventional procedures such as percutaneous coronary intervention (PCI), transcatheter aortic valve implantation (TAVI), pacemaker implantation (PI) and ablations has increased within the previous decade. Simultaneously, novel fluoroscopy mainframes enable lower radiation doses for patients and operators. Therefore, there is a need to update the existing diagnostic reference levels (DRLs) and propose new ones for common or recently introduced procedures. We sought to assess patient radiation doses in interventional cardiology in a large sample from seven hospitals across Finland between 2014 and 2016. Data were used to set updated national DRLs for coronary angiographies (kerma-air product (KAP) 30 Gycm2) and PCIs (KAP 75 cm2), and novel levels for PIs (KAP 3.5 Gycm2), atrial fibrillation ablation procedures (KAP 25 Gycm2) and TAVI (KAP 90 Gycm2). Tentative KAP values were set for implantations of cardiac resynchronization therapy devices (CRT, KAP 22 Gycm2), electrophysiological treatment of atrioventricular nodal re-entry tachycardia (6 Gycm2) and atrial flutter procedures (KAP 16 Gycm2). The values for TAVI and CRT device implantation are published for the first time on national level. Dose from image acquisition (cine) constitutes the major part of the total dose in coronary and atrial fibrillation ablation procedures. For TAVI, patient weight is a good predictor of patient dose.

Kokoelmat
  • Rinnakkaistallenteet [19207]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste