Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
Kaisu Rantakokko-Jalava; Kati Hokynar; Niina Hieta; Anniina Keskitalo; Pia Jokela; Anna Muotiala; T. Sakari Jokiranta; Rutta Kuusela; Hannu Sarkkinen; Janne Aittoniemi; Tytti Vuorinen; Antti J. Hakanen; Mirja Puolakkainen
Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
Kaisu Rantakokko-Jalava
Kati Hokynar
Niina Hieta
Anniina Keskitalo
Pia Jokela
Anna Muotiala
T. Sakari Jokiranta
Rutta Kuusela
Hannu Sarkkinen
Janne Aittoniemi
Tytti Vuorinen
Antti J. Hakanen
Mirja Puolakkainen
EUR CENTRE DIS PREVENTION & CONTROL
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2021042825804
https://urn.fi/URN:NBN:fi-fe2021042825804
Tiivistelmä
Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.