Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
Hannu Sarkkinen; Janne Aittoniemi; Anna Muotiala; Anniina Keskitalo; Niina Hieta; Tytti Vuorinen; Mirja Puolakkainen; Rutta Kuusela; T. Sakari Jokiranta; Antti J. Hakanen; Kaisu Rantakokko-Jalava; Pia Jokela; Kati Hokynar
Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019
Hannu Sarkkinen
Janne Aittoniemi
Anna Muotiala
Anniina Keskitalo
Niina Hieta
Tytti Vuorinen
Mirja Puolakkainen
Rutta Kuusela
T. Sakari Jokiranta
Antti J. Hakanen
Kaisu Rantakokko-Jalava
Pia Jokela
Kati Hokynar
EUR CENTRE DIS PREVENTION & CONTROL
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2021042825804
https://urn.fi/URN:NBN:fi-fe2021042825804
Tiivistelmä
Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.