Effect of respiratory motion correction and CT-based attenuation correction on dual-gated cardiac PET image quality and quantification

Saraste Antti; Klén Riku; Teuho Jarmo; Teräs Mika; Siekkinen Reetta; Lehtonen Eero; Schultz Jussi; Jafari Tadi Mojtaba

Effect of respiratory motion correction and CT-based attenuation correction on dual-gated cardiac PET image quality and quantification

Saraste Antti
Klén Riku
Teuho Jarmo
Teräs Mika
Siekkinen Reetta
Lehtonen Eero
Schultz Jussi
Jafari Tadi Mojtaba
Katso/Avaa
Publisher's pdf (1.129Mb)
Lataukset: 

Springer
doi:10.1007/s12350-021-02769-6
URI
https://link.springer.com/article/10.1007/s12350-021-02769-6
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2021093048586
Tiivistelmä

Background.

Dual-gating reduces respiratory and cardiac motion effects but increases noise. With motion correction, motion is minimized and image quality preserved. We applied motion correction to create end-diastolic respiratory motion corrected images from dual-gated images.

Methods.

[F-18]-fluorodeoxyglucose ([F-18]-FDG) PET images of 13 subjects were reconstructed with 4 methods: non-gated, dual-gated, motion corrected, and motion corrected with 4D-CT (MoCo-4D). Image quality was evaluated using standardized uptake values, contrast ratio, signal-to-noise ratio, coefficient of variation, and contrast-to-noise ratio. Motion minimization was evaluated using myocardial wall thickness.

Results.

MoCo-4D showed improvement for contrast ratio (2.83 vs 2.76), signal-to-noise ratio (27.5 vs 20.3) and contrast-to-noise ratio (14.5 vs 11.1) compared to dual-gating. The uptake difference between MoCo-4D and non-gated images was non-significant (P > .05) for the myocardium (2.06 vs 2.15 g/mL), but significant (P < .05) for the blood pool (.80 vs .86 g/mL). Non-gated images had the lowest coefficient of variation (27.3%), with significant increase for all other methods (31.6-32.5%). MoCo-4D showed smallest myocardial wall thickness (16.6 mm) with significant decrease compared to non-gated images (20.9 mm).

Conclusions.

End-diastolic respiratory motion correction and 4D-CT resulted in improved motion minimization and image quality over standard dual-gating.

Kokoelmat
  • Rinnakkaistallenteet [19207]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste