Analytical performance of point-of-care device QuikRead go in HbA1c analysis

Linko-Parvinen Anna; Dahlström Johanna; Pallari Hanna-Mari

Analytical performance of point-of-care device QuikRead go in HbA1c analysis

Linko-Parvinen Anna
Dahlström Johanna
Pallari Hanna-Mari
Katso/Avaa
8098-PB7-4165-R3.pdf (302.2Kb)
Lataukset: 

AME Publishing Company
doi:10.21037/jlpm-23-21
URI
https://doi.org/10.21037/jlpm-23-21
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082785697
Tiivistelmä

Background: Point-of-care (POC) analysers are crucial in acute care diagnostics. Due to easy access, they are also used in monitoring chronic diseases. However, use of POC devises in measurement of glycated haemoglobin (HbA1c) and monitoring diabetes mellitus is controversial.

Methods: We compared QuikRead go analyser’s HbA1c measurement performance to cobas c513 in a clinical laboratory with leftover venous K2-ethylenediaminetetraacetic acid (K2-EDTA) samples (n=151).

Results: The methods correlated well (r=0.984), but differed statistically significantly (P<0.0001) with an average negative bias of 2 mmol/mol for QuikRead go. The method deviation compared to cobas c513 was high, −20 to 11 mmol/mol (min–max), in HbA1c above 48 mmol/mol (n=98), and less, −5.5 to 1.4 mmol/mol, below HbA1c 48 mmol/mol (n=51).

Conclusions: This deviation in high level HbA1c should not be considered clinically acceptable. Deviating results can cause erroneous decisions regarding diabetes mellitus diagnosis and treatment. In lower HbA1c, the device showed relatively good performance. Concluding from this, the deviation is method-dependent and not related solely on preanalytical issues. These findings demonstrate the importance of understanding analytical variation when laboratory results and various methods are used in monitoring chronic diseases. Further studies with venous and finger-prick samples on QuikRead go analyser are needed to confirm the findings presented here.

Kokoelmat
  • Rinnakkaistallenteet [29335]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste