Feasibility of shortening scan duration of 18F-FDG myocardial metabolism imaging using a total-body PET/CT scanner

Zhang, Xiaochun; Xiang, Zeyin; Wang, Fanghu; Han, Chunlei; Zhang, Qing; Liu, Entao; Yuan, Hui; Jiang, Lei

Feasibility of shortening scan duration of 18F-FDG myocardial metabolism imaging using a total-body PET/CT scanner

Zhang, Xiaochun
Xiang, Zeyin
Wang, Fanghu
Han, Chunlei
Zhang, Qing
Liu, Entao
Yuan, Hui
Jiang, Lei
Katso/Avaa
s40658-024-00689-1.pdf (1.986Mb)
Lataukset: 

Springer Nature
doi:10.1186/s40658-024-00689-1
URI
https://ejnmmiphys.springeropen.com/articles/10.1186/s40658-024-00689-1
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082786168
Tiivistelmä

Purpose:

To evaluate 18F-FDG myocardial metabolism imaging (MMI) using a total-body PET/CT scanner and explore the feasible scan duration to guide the clinical practice.

Methods:

A retrospective analysis was conducted on 41 patients who underwent myocardial perfusion-metabolism imaging to assess myocardial viability. The patients underwent 18F-FDG MMI with a total-body PET/CT scanner using a list-mode for 600 s. PET data were trimmed and reconstructed to simulate images of 600-s, 300-s, 120-s, 60-s, and 30-s acquisition time (G600-G30). Images among different groups were subjectively evaluated using a 5-point Likert scale. Semi-quantitative evaluation was performed using standardized uptake value (SUV), myocardial to background activity ratio (M/B), signal to noise ratio (SNR), contrast to noise ratio (CNR), contrast ratio (CR), and coefficient of variation (CV). Myocardial viability analysis included indexes of Mismatch and Scar. G600 served as the reference.

Results:

Subjective visual evaluation indicated a decline in the scores of image quality with shortening scan duration. All the G600, G300, and G120 images were clinically acceptable (score ≥ 3), and their image quality scores were 4.9 ± 0.3, 4.8 ± 0.4, and 4.5 ± 0.8, respectively (P > 0.05). Moreover, as the scan duration reduced, the semi-quantitative parameters M/B, SNR, CNR, and CR decreased, while SUV and CV increased, and significant difference was observed in G300-G30 groups when comparing to G600 group (P < 0.05). For myocardial viability analysis of left ventricular and coronary segments, the Mismatch and Scar values of G300-G30 groups were almost identical to G600 group (ICC: 0.968-1.0, P < 0.001).

Conclusion:

Sufficient image quality for clinical diagnosis could be achieved at G120 for MMI using a total-body PET/CT scanner, while the image quality of G30 was acceptable for myocardial viability analysis.

Kokoelmat
  • Rinnakkaistallenteet [27094]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste