Evaluating Piezoelectric Ballistocardiography for Post-Surgical Heart Rate Monitoring

Seifizarei, Sepehr; Elnaggar, Ismail; Rikken, Lars; Anzanpour, Arman; Linckens, Stephan; Toffoli, Enrico; Vandewauw, Ine; Van De Poll; Marcel CG; Airola, Antti; Kaisti, Matti; De Kok, Margreet; Koivisto, Tero

Evaluating Piezoelectric Ballistocardiography for Post-Surgical Heart Rate Monitoring

Seifizarei, Sepehr
Elnaggar, Ismail
Rikken, Lars
Anzanpour, Arman
Linckens, Stephan
Toffoli, Enrico
Vandewauw, Ine
Van De Poll
Marcel CG
Airola, Antti
Kaisti, Matti
De Kok, Margreet
Koivisto, Tero
Katso/Avaa
CinC2024-309.pdf (3.680Mb)
Lataukset: 

doi:10.22489/cinc.2024.309
URI
https://doi.org/10.22489/cinc.2024.309
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082786879
Tiivistelmä

Ballistocardiography (BCG) is a non-invasive technique measuring heart and circulatory activity, proposed as an alternative to ECG for long-term heart rate monitoring. This study evaluates a custom multi-node piezoelectric BCG sensing mat for heart rate measurement compared to single-lead ECG in post-cardiac surgery patients. 14 patients (10 males, mean age 63.47 ± 9.35) postcoronary artery bypass (N=8) or heart valve repair (N=6) were enrolled at Maastricht University Medical Center. Patients in the cardiac ICU used a bed with a custom BCG sensor mat (10 PVDF-TrFE piezoelectric sensors, 90x50 cm²). BCG and ECG were recorded simultaneously for an average of 15 hours per patient. BCG heart rate values were extracted over ten-second epochs using a multinode rule-based algorithm. Performance was assessed using mean absolute error (MAE) comparing BCG and ECG heart rates. Heart rate values derived from the piezoelectric BCG mat resulted in a weighted mean MAE of 1.41 bpm with 66.14% coverage compared to ECG. Piezoelectric BCG shows promise as an accurate heart rate measurement method in post-cardiac surgery patients, offering a non-invasive alternative to ECG. These findings support further development of this technology for clinical use in remote heart rate monitoring.

Kokoelmat
  • Rinnakkaistallenteet [27094]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste