A Proof-of-Concept Study on Smart Gloves for Real-Time Chest Compression Performance Monitoring

Guridi Sofia; Mauranen Henry; Pouta Emmi; Semjonova Guna; Kahsay Desale Tewelde; Souza Leite Clayton; Rosio Riitta; Peltonen Laura-Maria; Tommila Miretta; Salantera Sanna

A Proof-of-Concept Study on Smart Gloves for Real-Time Chest Compression Performance Monitoring

Guridi Sofia
Mauranen Henry
Pouta Emmi
Semjonova Guna
Kahsay Desale Tewelde
Souza Leite Clayton
Rosio Riitta
Peltonen Laura-Maria
Tommila Miretta
Salantera Sanna
Katso/Avaa
A_Proof-of-Concept_Study_on_Smart_Gloves_for_.pdf (2.912Mb)
Lataukset: 

IEEE
doi:10.1109/ACCESS.2024.3361663
URI
https://ieeexplore.ieee.org/document/10418938
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082787103
Tiivistelmä
Correctly performed Cardiopulmonary Resuscitation (CPR) is a critical element in preventing deaths caused by cardiac arrest (CA). To improve the outcomes and quality of CPR, stand-alone devices that monitor the performance and provide feedback have been developed. However, these devices have multiple limitations due to their rigidity and stiffness. Furthermore, most of the devices do not account for complete chest recoil as a metric of quality CPR, reducing the quality of compressions. To overcome these limitations, this study proposes smart gloves equipped with e-textiles-based pressure sensors and inertial measurement units (IMUs) to monitor the quality adult CPR in real-time. The prototype development combined data-driven design and Research Through Design (RtD) methods, taking into account not only the accuracy but also the usability of the smart gloves. A preliminary study with nine participants performing CPR on a doll was conducted to evaluate the accuracy and wearability of the smart gloves. Study results show that the smart gloves accurately detect chest compression parameters, including compression depth, compression rate, chest recoil and interruption between compressions based on the intelligent fusion of pressure sensors and IMUs. In addition, the newly developed smart gloves are lightweight, hand adaptable, and easily replicable as an alternative for hard case devices. The design methods used in this study can be applied to design other accessible and comfortable wearable devices in healthcare settings. © 2013 IEEE.
Kokoelmat
  • Rinnakkaistallenteet [27094]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste