A Comparison Between a Resuscitation Glove and Standard Manual Compressions on the Quality of Cardiovascular Resuscitation

Kahsay, Desale Tewelde; Tommila, Miretta; Peltonen, Laura-Maria; Löyttyniemi, Eliisa; Xiao, Yu; Mauranen, Henry; Salanterä, Sanna

A Comparison Between a Resuscitation Glove and Standard Manual Compressions on the Quality of Cardiovascular Resuscitation

Kahsay, Desale Tewelde
Tommila, Miretta
Peltonen, Laura-Maria
Löyttyniemi, Eliisa
Xiao, Yu
Mauranen, Henry
Salanterä, Sanna
Katso/Avaa
a_comparison_between_a_resuscitation_glove_and.285.pdf (653.2Kb)
Lataukset: 

Ovid Technologies (Wolters Kluwer Health)
doi:10.1097/JCN.0000000000001206
URI
https://doi.org/10.1097/jcn.0000000000001206
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082792833
Tiivistelmä

Background 

Several audiovisual feedback (AVF) devices have been developed to monitor chest compression quality during cardiopulmonary resuscitation (CPR). However, most marketed stand-alone AVF devices are inflexible and rigid, causing discomfort and sometimes pain to the rescuers' hands.

Objective 

The objective of this study was to evaluate the effectiveness and usability of a newly developed soft and flexible resuscitation glove designed to improve the quality of chest compressions during CPR.

Methods 

We conducted a manikin-based randomized crossover study to compare the effectiveness of a newly developed AVF device (ResuGlove CPR Group) and standard CPR (Standard CPR Group) in improving the quality of chest compressions in simulated cardiac arrest cases. The usability of the newly developed ResuGlove was assessed using a System Usability Scale questionnaire.

Results 

There were no significant differences in compression depth (mean, 53.69 vs 53.28; P = .70) and compression rate (mean, 111.48 vs 113.38; P = .23) between the ResuGlove CPR and Standard CPR groups. However, the group using ResuGlove had a higher percentage of complete chest releases between compressions (P = .008). Furthermore, the ResuGlove CPR Group had a significantly higher percentage of participants who performed chest compressions with adequate compression depth (82.8% vs 41.4%, P = .001) and compression rate (96.6% vs 72.4%, P = .012) compared with the Standard CPR Group. The ResuGlove usability score was calculated to be 70.4.

Conclusions 

The newly developed ResuGlove significantly improved the quality of certain chest compression parameters, and the device's usability score was within the acceptable range.

Kokoelmat
  • Rinnakkaistallenteet [29337]

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste
 

 

Tämä kokoelma

JulkaisuajatTekijätNimekkeetAsiasanatTiedekuntaLaitosOppiaineYhteisöt ja kokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy

Turun yliopiston kirjasto | Turun yliopisto
julkaisut@utu.fi | Tietosuoja | Saavutettavuusseloste