CLinical EValuation of WEB 17 device in intracranial aneuRysms (CLEVER): 1-year effectiveness results for ruptured and unruptured aneurysms
| dc.contributor.author | Spelle, Laurent | |
| dc.contributor.author | Costalat, Vincent | |
| dc.contributor.author | Caroff, Jildaz | |
| dc.contributor.author | Wodarg, Fritz | |
| dc.contributor.author | Fischer, Sebastian | |
| dc.contributor.author | Herbreteau, Denis | |
| dc.contributor.author | Möhlenbruch, Markus A | |
| dc.contributor.author | Januel, Anne-Christine | |
| dc.contributor.author | Papagiannaki, Chrysanthi | |
| dc.contributor.author | Klisch, Joachim | |
| dc.contributor.author | Numminen, Jussi | |
| dc.contributor.author | Rautio, Riitta | |
| dc.contributor.author | Berlis, Ansgar | |
| dc.contributor.author | Mihalea, Cristian | |
| dc.contributor.author | Chalumeau, Vanessa | |
| dc.contributor.author | Downer, Jonathan | |
| dc.contributor.author | Cortese, Jonathan | |
| dc.contributor.author | Ikka, Leon | |
| dc.contributor.author | Gallas, Sophie | |
| dc.contributor.author | Bester, Maxim | |
| dc.contributor.author | Liebig, Thomas | |
| dc.contributor.author | Velasco, Stéphane | |
| dc.contributor.author | Grimaldi, Lamiae | |
| dc.contributor.author | Byrne, James | |
| dc.contributor.author | Szikora, Istvan | |
| dc.contributor.author | Pierot, Laurent | |
| dc.contributor.author | Cognard, Christophe | |
| dc.contributor.organization | fi=lääketieteellinen tiedekunta|en=Faculty of Medicine| | |
| dc.contributor.organization | fi=tyks, vsshp|en=tyks, varha| | |
| dc.contributor.organization-code | 1.2.246.10.2458963.20.13290506867 | |
| dc.converis.publication-id | 457219695 | |
| dc.converis.url | https://research.utu.fi/converis/portal/Publication/457219695 | |
| dc.date.accessioned | 2025-08-28T00:50:54Z | |
| dc.date.available | 2025-08-28T00:50:54Z | |
| dc.description.abstract | <p><strong>Background: </strong>The Woven EndoBridge (WEB) device is designed to treat wide-necked bifurcation aneurysms. The WEB 17 is the latest iteration and can be delivered through a 0.017″ microcatheter. The CLEVER study demonstrated that WEB 17 is safe and effective for providing protection against bleeding or rebleeding at 1 month and 1 year.</p><p><strong>Objective: </strong>To evaluate angiographic stability at 1 year.</p><p><strong>Methods: </strong>The CLEVER study was a prospective multicenter study conducted in 17 European centers, involved 163 subjects, comprising 60 ruptured and 103 unruptured aneurysms. Independent assessment of 1-year follow-up imaging was incorporated into the study design.</p><p><strong>Results: </strong>Aneurysm diameters ranged from 2.0 to 9.2 mm, with 95.7% being broad-based (dome-to-neck ratio <2). Follow-up imaging at 1 year was completed for 146 out of 163 subjects (89.6%) and evaluated by an independent core laboratory. The primary efficacy endpoint of adequate occlusion without re-treatment at 1 year was achieved for 120 (82.2%) of all subjects. At 1 year, the adequate occlusion rate was 86.5% for ruptured aneurysms (73.1% complete occlusion) and 82.4% for unruptured aneurysms (57.1% complete occlusion). The overall re-treatment rate at 1 year was 2.6% (4/152), with 3.1% (3/97) for unruptured aneurysms and 1.8% (1/55) for ruptured aneurysms CONCLUSION: Delivery of the WEB 17 via 0.017 inch catheters represents a significant evolution of the WEB design. The results of CLEVER presented here demonstrate that it maintains the same efficacy as previous generations of WEB.</p> | |
| dc.identifier.eissn | 1759-8486 | |
| dc.identifier.jour-issn | 1759-8478 | |
| dc.identifier.olddbid | 206531 | |
| dc.identifier.oldhandle | 10024/189558 | |
| dc.identifier.uri | https://www.utupub.fi/handle/11111/47079 | |
| dc.identifier.url | https://jnis.bmj.com/content/early/2024/07/10/jnis-2024-021918 | |
| dc.identifier.urn | URN:NBN:fi-fe2025082787390 | |
| dc.language.iso | en | |
| dc.okm.affiliatedauthor | Rautio, Riitta | |
| dc.okm.affiliatedauthor | Dataimport, tyks, vsshp | |
| dc.okm.discipline | 3112 Neurosciences | en_GB |
| dc.okm.discipline | 3126 Surgery, anesthesiology, intensive care, radiology | en_GB |
| dc.okm.discipline | 3112 Neurotieteet | fi_FI |
| dc.okm.discipline | 3126 Kirurgia, anestesiologia, tehohoito, radiologia | fi_FI |
| dc.okm.internationalcopublication | international co-publication | |
| dc.okm.internationality | International publication | |
| dc.okm.type | A1 ScientificArticle | |
| dc.publisher | BMJ Publishing Group | |
| dc.publisher.country | United Kingdom | en_GB |
| dc.publisher.country | Britannia | fi_FI |
| dc.publisher.country-code | GB | |
| dc.relation.doi | 10.1136/jnis-2024-021918 | |
| dc.relation.ispartofjournal | Journal of NeuroInterventional Surgery | |
| dc.source.identifier | https://www.utupub.fi/handle/10024/189558 | |
| dc.title | CLinical EValuation of WEB 17 device in intracranial aneuRysms (CLEVER): 1-year effectiveness results for ruptured and unruptured aneurysms | |
| dc.year.issued | 2024 |
Tiedostot