Bioimpedance spinal needle provides high success and low complication rate in lumbar punctures of pediatric patients with acute lymphoblastic leukemia

dc.contributor.authorLångström Satu
dc.contributor.authorHuurre Anu
dc.contributor.authorKari Juho
dc.contributor.authorLohi Olli
dc.contributor.authorSievänen Harri
dc.contributor.authorPalmu Sauli
dc.contributor.organizationfi=lastentautioppi|en=Paediatrics and Adolescent Medicine|
dc.contributor.organizationfi=tyks, vsshp|en=tyks, varha|
dc.contributor.organization-code1.2.246.10.2458963.20.40612039509
dc.converis.publication-id175461372
dc.converis.urlhttps://research.utu.fi/converis/portal/Publication/175461372
dc.date.accessioned2022-10-28T13:42:10Z
dc.date.available2022-10-28T13:42:10Z
dc.description.abstract<p>In this prospective single-arm study of 50 pediatric patients with acute lymphoblastic leukemia (ALL), we evaluated the clinical performance of a novel bioimpedance spinal needle system in 152 intrathecal treatment lumbar punctures (LP) of these patients. The system detects in real-time when the needle tip reaches the cerebrospinal fluid (CSF) in the spinal canal. The success was defined as getting a CSF sample and/or administering the intrathecal treatment with one needle insertion. Incidence of traumatic LP (TLP) was defined as >= 10 erythrocytes/mu L of CSF. Post-procedural complications were monitored with a one-week diary and one-month register follow-up. The success of the first attempt was 79.5%, with the CSF detection sensitivity of 86.1%. The incidence of TLP was 17.3%. A successful first attempt was associated with a significantly lower incidence of TLP (10% vs 40%, p = 0.0015). During the week after the procedure, the incidence of post-dural puncture headache was 6%. During the follow-up, no major complications were observed. In conclusion, the novel bioimpedance spinal needle system achieved a high success rate and low incidence of TLP and other complications in pediatric patients with ALL in a real-world clinical setting, indicating clinical utility for this system in pediatric hemato-oncology.<br></p>
dc.identifier.eissn2045-2322
dc.identifier.jour-issn2045-2322
dc.identifier.olddbid183736
dc.identifier.oldhandle10024/166830
dc.identifier.urihttps://www.utupub.fi/handle/11111/41059
dc.identifier.urlhttps://doi.org/10.1038/s41598-022-10915-4
dc.identifier.urnURN:NBN:fi-fe2022081154619
dc.language.isoen
dc.okm.affiliatedauthorHuurre, Anu
dc.okm.affiliatedauthorDataimport, tyks, vsshp
dc.okm.discipline3122 Cancersen_GB
dc.okm.discipline3123 Gynaecology and paediatricsen_GB
dc.okm.discipline3122 Syöpätauditfi_FI
dc.okm.discipline3123 Naisten- ja lastentauditfi_FI
dc.okm.internationalcopublicationnot an international co-publication
dc.okm.internationalityInternational publication
dc.okm.typeA1 ScientificArticle
dc.publisherNature Publishing Group
dc.publisher.countryUnited Kingdomen_GB
dc.publisher.countryBritanniafi_FI
dc.publisher.country-codeGB
dc.relation.articlenumber6799
dc.relation.doi10.1038/s41598-022-10915-4
dc.relation.ispartofjournalScientific Reports
dc.relation.issue1
dc.relation.volume12
dc.source.identifierhttps://www.utupub.fi/handle/10024/166830
dc.titleBioimpedance spinal needle provides high success and low complication rate in lumbar punctures of pediatric patients with acute lymphoblastic leukemia
dc.year.issued2022

Tiedostot

Näytetään 1 - 1 / 1
Ladataan...
Name:
s41598-022-10915-4.pdf
Size:
1.13 MB
Format:
Adobe Portable Document Format