Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation

dc.contributor.authorMatti Kaisti
dc.contributor.authorTuukka Panula
dc.contributor.authorJoni Leppänen
dc.contributor.authorRisto Punkkinen
dc.contributor.authorMojtaba Jafari Tadi
dc.contributor.authorTuija Vasankari
dc.contributor.authorSamuli Jaakkola
dc.contributor.authorTuomas Kiviniemi
dc.contributor.authorJuhani Airaksinen
dc.contributor.authorPekka Kostiainen
dc.contributor.authorUlf Meriheinä
dc.contributor.authorTero Koivisto
dc.contributor.authorMikko Pänkäälä
dc.contributor.organizationfi=sisätautioppi|en=Internal Medicine|
dc.contributor.organizationfi=sulautettu elektroniikka|en=Embedded Electronics|
dc.contributor.organizationfi=terveysteknologia|en=Health Technology|
dc.contributor.organizationfi=tyks, vsshp|en=tyks, varha|
dc.contributor.organization-code1.2.246.10.2458963.20.20754768032
dc.contributor.organization-code1.2.246.10.2458963.20.28696315432
dc.contributor.organization-code1.2.246.10.2458963.20.40502528769
dc.converis.publication-id40651802
dc.converis.urlhttps://research.utu.fi/converis/portal/Publication/40651802
dc.date.accessioned2022-10-28T13:16:12Z
dc.date.available2022-10-28T13:16:12Z
dc.description.abstractThere is an unmet clinical need for a low cost and easy to use wearable devices for continuous cardiovascular health monitoring. A flexible and wearable wristband, based on microelectromechanical sensor (MEMS) elements array was developed to support this need. The performance of the device in cardiovascular monitoring was investigated by (i) comparing the arterial pressure waveform recordings to the gold standard, invasive catheter recording (n = 18), (ii) analyzing the ability to detect irregularities of the rhythm (n = 7), and (iii) measuring the heartrate monitoring accuracy (n = 31). Arterial waveforms carry important physiological information and the comparison study revealed that the recordings made with the wearable device and with the gold standard device resulted in almost identical (r = 0.9-0.99) pulse waveforms. The device can measure the heart rhythm and possible irregularities in it. A clustering analysis demonstrates a perfect classification accuracy between atrial fibrillation (AF) and sinus rhythm. The heartrate monitoring study showed near perfect beat-to-beat accuracy (sensitivity = 99.1%, precision = 100%) on healthy subjects. In contrast, beat-to-beat detection from coronary artery disease patients was challenging, but the averaged heartrate was extracted successfully (95% CI: -1.2 to 1.1 bpm). In conclusion, the results indicate that the device could be useful in remote monitoring of cardiovascular diseases and personalized medicine.
dc.identifier.eissn2398-6352
dc.identifier.jour-issn2398-6352
dc.identifier.olddbid180926
dc.identifier.oldhandle10024/164020
dc.identifier.urihttps://www.utupub.fi/handle/11111/29256
dc.identifier.urnURN:NBN:fi-fe2021042822132
dc.language.isoen
dc.okm.affiliatedauthorKaisti, Matti
dc.okm.affiliatedauthorPanula, Tuukka
dc.okm.affiliatedauthorPunkkinen, Risto
dc.okm.affiliatedauthorJafari Tadi, Mojtaba
dc.okm.affiliatedauthorJaakkola, Samuli
dc.okm.affiliatedauthorKiviniemi, Tuomas
dc.okm.affiliatedauthorAiraksinen, Juhani
dc.okm.affiliatedauthorKoivisto, Tero
dc.okm.affiliatedauthorPänkäälä, Mikko
dc.okm.affiliatedauthorDataimport, tyks, vsshp
dc.okm.discipline217 Medical engineeringen_GB
dc.okm.discipline3121 Internal medicineen_GB
dc.okm.discipline217 Lääketieteen tekniikkafi_FI
dc.okm.discipline3121 Sisätauditfi_FI
dc.okm.internationalcopublicationinternational co-publication
dc.okm.internationalityInternational publication
dc.okm.typeA1 ScientificArticle
dc.publisherNATURE PUBLISHING GROUP
dc.publisher.countryUnited Kingdomen_GB
dc.publisher.countryBritanniafi_FI
dc.publisher.country-codeGB
dc.relation.articlenumber39
dc.relation.doi10.1038/s41746-019-0117-x
dc.relation.ispartofjournalnpj Digital Medicine
dc.relation.volume2
dc.source.identifierhttps://www.utupub.fi/handle/10024/164020
dc.titleClinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation
dc.year.issued2019

Tiedostot

Näytetään 1 - 1 / 1
Ladataan...
Name:
s41746-019-0117-x.pdf
Size:
2.01 MB
Format:
Adobe Portable Document Format
Description:
Publisher's PDF (CC BY)