Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring
Panula Tuukka; Sirkiä Jukka-Pekka; Koivisto Tero; Pänkäälä Mikko; Niiranen Teemu; Kantola Ilkka; Kaisti Matti
Development and clinical validation of a miniaturized finger probe for bedside hemodynamic monitoring
Panula Tuukka
Sirkiä Jukka-Pekka
Koivisto Tero
Pänkäälä Mikko
Niiranen Teemu
Kantola Ilkka
Kaisti Matti
Cell Press
Julkaisun pysyvä osoite on:
https://urn.fi/URN:NBN:fi-fe2025082792101
https://urn.fi/URN:NBN:fi-fe2025082792101
Tiivistelmä
Our aim is to develop a blood pressure (BP) measurement technology that could be integrated into a finger-worn pulse oximeter, eliminating the need for a brachial cuff. We present a miniature cuffless tonometric finger probe system that uses the oscillometric method to measure BP. Our approach uses a motorized press that is used to apply pressure to the fingertip to measure BP. We verified the functionality of the device in a clinical trial (n = 43) resulting in systolic and diastolic pressures ((mean ± SD) mmHg) of (-3.5 ± 8.4) mmHg and (-4.0 ± 4.4) mmHg, respectively. Comparison was made with manual auscultation (n = 26) and automated cuff oscillometry (n = 18). In addition to BP, we demonstrated the ability of the device to assess arterial stiffness (n = 18) and detect atrial fibrillation (n = 6). We were able to introduce a sufficiently small device that could be used for convenient ambulatory measurements with minimal discomfort.